Mouse TNF-alpha ELISA Kit

For detection and measurement of mouse tumor necrosis factor-alpha
概要
The Mouse Tumor Necrosis Factor-alpha (TNF-α) ELISA Kit is designed for the quantitative detection and measurement of mouse TNF-α in biological fluids such as serum, plasma, and cell culture supernatants. TNF-α is primarily produced by activated T cells and macrophages in response to inflammation or infection, but is also produced by B cells, NK cells, mast cells, neutrophils, dendritic cells, microglia, endothelial cells, smooth muscle cells, cardiomyocytes, and fibroblasts. TNF-α is a pro-inflammatory cytokine and stimulates anti-tumor responses in vitro.

The assay is based on the sandwich ELISA method, in which samples are added to ELISA strip plates pre-coated with capture antibodies specific for the cytokine. The captured cytokine is detected by addition of a biotinylated detection antibody, followed by streptavidin-horseradish peroxidase, which binds the biotinylated antibody. Addition of the chromogenic enzyme substrate 3,3’,5,5’ tetramethylbenzidine (TMB) results in a colored product with an intensity directly proportional to the concentration of cytokine in the sample. The concentration of the cytokine is determined by comparison to a serial dilution of the cytokine standard analyzed in parallel.
Components
  • Mouse TNF-alpha ELISA Kit, 2 plates (Catalog #02030)
    • Mouse TNF-α ELISA Plate, 2x
    • Mouse TNF-α Standard, 1 vial
    • Mouse TNF-α Biotinylated Detection Antibody, 30 µL
    • SA-HRP, 30 µL
    • Standard Reconstitution Buffer B, 1 mL
    • Wash Buffer (20X), 120 mL
    • ELISA Diluent, 120 mL
    • SA-HRP Diluent, 25 mL
    • TMB Substrate, 25 mL
    • Stop Solution, 25 mL
    • Adhesive Plate Covers, 6x
 
  • Mouse TNF-alpha ELISA Kit, 10 plates (Catalog #02031)
    • Mouse TNF-α  ELISA Plate, 10x
    • Mouse TNF-α Standard, 1 vial
    • Mouse TNF-α Biotinylated Detection Antibody, 130 µL
    • SA-HRP, 130 µL
    • Standard Reconstitution Buffer B, 1 mL
    • Wash Buffer (20X), 5 x 120 mL
    • ELISA Diluent, 3 x 120 mL
    • SA-HRP Diluent, 120 mL
    • TMB Substrate, 120 mL
    • Stop Solution, 120 mL
    • Adhesive Plate Covers, 30x
Subtype
Complete Kits
Cell Type
B Cells, Cardiomyocytes, PSC-Derived, Dendritic Cells, Endothelial Cells, Granulocytes and Subsets, Macrophages, Mouse Embryonic Fibroblasts, Myogenic Stem and Progenitor Cells, NK Cells, T Cells
Species
Mouse
Application
ELISA
Area of Interest
Immunology, Stem Cell Biology
数据及文献

Data

Representative Standard Curve

• Reportable range: 15.6 - 2000 pg/mL. This is the concentration range in which measurement of the analyte can be done with the highest precision, accuracy, and linearity.
• Sensitivity: The limit of detection of this assay is 5 pg/mL. This is the analyte concentration with absorbance two standard deviations higher than the zero standard.
• Accuracy: The analyte standard of this ELISA was calibrated against NIBSC international standard 88/532.
• Recovery: A mid-curve recovery of 84 - 86% was determined by spiking defined amounts of analyte standard into serum or plasma samples in repeated experiments.
• Precision: The intra-assay precision of this assay is 2.4% (CV). The Inter-assay precision of this assay is 3.7% (CV).

* National Institute of Biological Standards and Control, Potters Bar, Hertfordshire EN6 3QG, UK.

Top